HG PRO LN – PEM Hydrogen Generator

Main Applications:

  • Carrier gas for GC and GS-MS
  • GC-FID / GC-NPD / GC-FPD / THA
  • ICP-MS collision gas
  • Refilling metal hydride tanks for use with fuel cells
  • CVD application for diamond production
HG PRO LN – PEM Hydrogen Generator

Product Features:

  • New long life multilayer PEM cell
  • High flow rate up to 1500 cc/min
  • High pressure up to 16 bar (232 psi)
  • N2 content < 1 ppm
  • RS 232/485 and USB standard
  • LAN optional
  • Auto refill standard
  • High purity: 99.99999%
  • Dryer with no maintenance
  • Safety routine and self monitoring
  • Reduced foot print
  • High pressure GLS electronically controlled (patented)

The HydroGen series HG PRO LN is able to generate up to 1500 cc/min of Hydrogen with purity better than 99.99999%, a pressure of up to 16 bar (232 psi), and an N2 content < 1 ppm.

The HG PRO includes a new long-life multilayer electrolytic cell with a polymeric membrane (PEM) using distilled water. No acid or alkaline solutions are used.

The innovative gas drying system (two-column drying PSA system with automatic regeneration) ensures the maximum grade of hydrogen purity, it is completely maintenance-free and allows continuous operation, 24 hours a day.

The onboard CPU automatically checks for internal leaks and constantly controls the operating parameters to guarantee full safety. Up to 20 units can be connected in parallel using an external control box.

The touch-screen LCD interface provides simple and user-friendly management of all functions on the unit.


Technical Specifications:

General data
Electrolytic cellLong life Multilayer PEM technology
Drying systemPSA (Pressure Swing Adsorption)
H2 purity>99.99999% *1
Outlet pressure12 bars (174 psi) / 16 bars (232 psi) (optional)
N2 content< 1 ppm
Dimensions (W x D x H)27 x 44 x 41 cm
Communication
RS-232 / USB / RS-485Standard
Digital I/OStandard
LANOptional
Software functions
Parallel modeReady on board
Automatic tank refillStandard
Fill Hydride canister functionStandard
Water
Quality RequiredDeionized, ASTM II, <0.1uS
Supply pressure (min, Max)0.1 bar (1.45 psi), 0.5 bar (7.25 psi)
Supply flow rate (min, Max)0.2 l/min, 1.5 l/min
Internal tank capacity3 L
External tank capacity5 or 10 L (optional)
Electrical data
Type of connectionIEC 320-C14
Power supply voltage100-240 Vac (±10%) 50/60 Hz
Fuse rating (5x20mm)6.3A (250 Vac – T)
Connections
Hydrogen outlet1/8” compression fitting
WaterQuick release push-in fitting
Operating/storage conditions
Working Temperature5-40°C (41-104°F) *2
Storage Temperature1-50°C (34-122°F)
Humidity (max, non condensing)70%
Noise< 35dB(A)
IP ratingIP20
Pollution degree rating2 (with no aromatic compounds)
Altitude< 2000m
*1 referred to O2, dew point <-80°C (< -112°F), Atmospheric Dew Point (ADP)
*2 with temperature > 35°C (> 95°F) the H2 purity could be less
Model: HG PRO LN700900110013001500
H2 flow-rate (Max)700900110013001500
Net weight (no water in tank)23.5 Kg26 Kg26 Kg26 Kg27 Kg
Nominal power (Max)640W640W640W640W750W